Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 31: (1 year) CAL VER /5520 INSTRUMENT: Fluke 33: (1 year) CAL VER /5520 DATE: 30-Oct-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 190 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Source: # Fluke 31/33 Instruction Manual (PN 896118 Sep. 1991). # Fluke 5500A/COIL Instruction Sheet (PN 600616 Aug. 1995) # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- R P F W 1.003 ASK+ K 1.004 HEAD EQUIPMENT SETUP 1.005 HEAD {RMS AMPS TESTS} 1.006 DISP Turn the UUT on. 1.007 DISP Connect the 5500A/COIL to the 5520A 20A & AUX Lo 1.007 DISP terminals. 1.007 DISP 1.007 DISP Clamp the meter around the toroid center conductor 1.007 DISP bundle, positioning at the alignment marks. 1.008 HEAD RMS AMPS TESTS: {400A Range} 1.009 5520 400.0A 60H SI HC S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 1.010 ACC 400.0A 8.2U 1.011 MEMI Enter UUT reading in Amps: 1.012 MEME 1.013 MEMC 400 A 8.2U 60H 2.001 5520 200.0A 60H SI HC S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 2.002 ACC 200.0A 4.2U 2.003 MEMI Enter UUT reading in Amps: 2.004 MEME 2.005 MEMC 400 A 4.2U 60H 3.001 JMPT 14.001 Fluke 31: (1 year) CAL VER /5520 3.002 HEAD {PEAK AMPS TESTS} 3.003 DISP Press the SMOOTH/CREST button twice to select PEAK. 3.004 HEAD PEAK AMPS TESTS: {400A Range} 3.005 5520 565.8Ap 60H SI HC S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 3.006 ACC 565.8Ap 17.2U 3.007 MEMI Enter UUT reading in Amps peak: 3.008 MEME 3.009 MEMC 400 Ap 17.2U 60H 4.001 5520 282.85Ap 60H SI HC S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 4.002 ACC 282.85Ap 8.75U 4.003 MEMI Enter UUT reading in Amps peak: 4.004 MEME 4.005 MEMC 400 Ap 8.75U 60H 5.001 DISP Connect the 5500A/COIL to the 5520A AUX Hi & Lo 5.001 DISP terminals. 5.001 DISP 5.001 DISP Clamp the meter around the toroid center conductor 5.001 DISP bundle, positioning at the alignment marks. 5.002 5520 141.45Ap 60H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 5.003 ACC 141.45Ap 4.55U 5.004 MEMI Enter UUT reading in Amps peak: 5.005 MEME 5.006 MEMC 400 Ap 4.55U 60H 6.001 5520 113.1Ap 50H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 6.002 ACC 113.1Ap 3.7U 6.003 MEMI Enter UUT reading in Amps peak: 6.004 MEME 6.005 MEMC 400 Ap 3.7U 50H 7.001 HEAD PEAK AMPS TESTS: {40A Range} 7.002 5520 56.6Ap 50H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 7.003 ACC 56.6Ap 2.0U 7.004 MEMI Enter UUT reading in Amps peak: 7.005 MEME 7.006 MEMC 40 Ap 2.0U 50H 8.001 5520 28.285Ap 400H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 8.002 ACC 28.285Ap 4.55U 8.003 MEMI Enter UUT reading in Amps peak: 8.004 MEME 8.005 MEMC 40 Ap 4.55U 400H 9.001 DISP Disconnect the 5500A/COIL from the 5520A. 9.002 DISP Loop a piece of wire of sufficient gauge from the 9.002 DISP 20A OUTPUT to the AUX OUTPUT LO. 9.002 DISP 9.002 DISP Clamp the meter around the loop positioning at the 9.002 DISP alignment marks. 9.003 5520 40 14.145Ap 0.865U 400H SI HC L2W 10.001 5520 40 11.31Ap 0.64U 60H SI HC L2W 11.001 5520 40 5.66Ap 0.47U 60H SI HC L2W 12.001 DISP Loop a piece of wire of sufficient gauge from the 12.001 DISP AUX OUTPUT HI to the AUX OUTPUT LO. 12.001 DISP 12.001 DISP Clamp the meter around the loop positioning at the 12.001 DISP alignment marks. 12.002 5520 40 2.825Ap 0.385U 50H SI L2W 13.001 5520 40 1.415Ap 0.345U 50H SI L2W 14.001 HEAD {RMS AMPS TESTS} 14.002 JMPT 14.005 Fluke 31: (1 year) CAL VER /5520 14.003 DISP Connect the 5500A/COIL to the 5520A AUX Hi & Lo 14.003 DISP terminals. 14.003 DISP 14.003 DISP Clamp the meter around the toroid center conductor 14.003 DISP bundle, positioning at the alignment marks. 14.004 DISP Press the SMOOTH/CREST button to select RMS. 14.005 HEAD RMS AMPS TESTS: {400A Range} 14.006 5520 100.0A 60H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 14.007 ACC 100.0A 2.2U 14.008 MEMI Enter UUT reading in Amps: 14.009 MEME 14.010 MEMC 400 A 2.2U 60H 15.001 5520 80.0A 50H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 15.002 ACC 80.0A 1.8U 15.003 MEMI Enter UUT reading in Amps: 15.004 MEME 15.005 MEMC 400 A 1.8U 50H 16.001 HEAD RMS AMPS TESTS: {40A Range} 16.002 5520 40.0A 50H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 16.003 ACC 40.0A 1.0U 16.004 MEMI Enter UUT reading in Amps: 16.005 MEME 16.006 MEMC 40 A 1.0U 50H 17.001 5520 20.00A 400H SI S LT5 # Need 5500A/COIL accuracy + contribution from UUT here. 17.002 ACC 20.00A 1.10U 17.003 MEMI Enter UUT reading in Amps: 17.004 MEME 17.005 MEMC 40 A 1.10U 400H 18.001 DISP Disconnect the 5500A/COIL from the 5520A. 18.002 DISP Loop a piece of wire of sufficient gauge from the 18.002 DISP 5520A 20A OUTPUT to the AUX OUTPUT LO. 18.002 DISP 18.002 DISP Clamp the meter around the loop positioning at the 18.002 DISP alignment marks. 18.003 5520 10.00A 0.70U 400H SI HC L2W 19.001 5520 8.00A 0.36U 60H SI HC L2W 20.001 5520 4.00A 0.24U 60H SI HC L2W 21.001 DISP Loop a piece of wire of sufficient gauge from the 21.001 DISP 5520A AUX OUTPUT HI to the AUX OUTPUT LO. 21.001 DISP 21.001 DISP Clamp the meter around the loop positioning at the 21.001 DISP alignment marks. 21.002 5520 2.00A 0.24U 50H SI L2W 22.001 5520 1.00A 0.22U 50H SI L2W 23.001 HEAD {FREQUENCY TESTS} 23.002 DISP Press the Hz button. 23.003 HEAD FREQUENCY TESTS: 40A Range 23.004 DISP Loop a piece of wire of sufficient gauge from the 23.004 DISP 5520A 20A OUTPUT to the AUX OUTPUT LO. 23.004 DISP 23.004 DISP Clamp the meter around the loop positioning at the 23.004 DISP alignment marks. 23.005 5520 400.0H 1.1U 10A SI HC L2W 24.001 5520 60.0H 0.4U 4A SI HC L2W 25.001 DISP Loop a piece of wire of sufficient gauge from the 25.001 DISP 5520A AUX OUTPUT HI to the AUX OUTPUT LO. 25.001 DISP 25.001 DISP Clamp the meter around the loop positioning at the 25.001 DISP alignment marks. 25.002 5520 50.0H 0.4U 1A SI L2W 26.001 END